Code: DM-53D
20L
|
|
INTENDED USE
(For In Vitro Diagnostic Use Only)
DM-53D Diluent is used by Mindray BC 5380 hematological analyzer as a whole blood sample DILUTING and SHEATH reagent.
SUMMARY AND PRINCIPLE
A sample volume of a whole blood specimen is aspirated into the analyzer where the sample is segmented into 4 portions in a shear valve. Two of these portions are diluted with DM-53D Diluent. To one of the dilutions a lysing reagent (DM-53LH Lyse) is added for the measurement of hemoglobin (HGB). The other dilution is introduced into impedance particle analyzer where the red blood cell count (RBC) and the thrombocyte count (PLT) is measured. DM-53D Diluent is also used by the instrument as a sheath fluid in the flow cell where the white blood cell count (WBC), lymphocyte count (LYM), monocyte cell count (MONO), eosinophil cell count (EO), basophil cell count (BASO), neutrophil cell count (NEU) measurement is performed. Consult your specific instrument Operator?s Manual for additional information with respect to procedures and principles for whole blood hematological analysis.