Code: 12304
|
|
INTENDED USE
(For In Vitro Diagnostic Use Only)
Dialyse-CD 3.0 is used by the Cell-Dyn 3000 hematological analyzer(s) as a hemolysing reagent for the assay of hemoglobin.
SUMMARY AND PRINCIPLE
A sample volume of a whole blood specimen is aspirated into the analyzer where a portion of it is automatically diluted with Diaton-CD 3/3.5 Diff. A portion of this dilution of the sample is then introduced into impedance cell where the red blood cell count (RBC) and the thrombocyte count (PLT) is measured. Another portion of the aspirated blood sample is also diluted and a hemoglobin lyse reagent (Dialyse-CD 3.0) is added for hemoglobin (HGB) measurement. A third portion of the aspirated blood sample is diluted with the sheath reagent (Diasheath-CD 3/3.5) for white blood cell count (WBC), lymphocyte count (LYM), monocyte cell count (MONO), eosinophil cell count (EO), basophil cell count (BASO), neutrophil cell count (NEU) measurement in the flow cytometer. Consult your specific instrument Operators Manual for additional information with respect to procedures and principles for whole blood hematological analysis.